Technical specifications for passive version
PRESSURE MEASURING RANGE (MH2O)
|
10 ... 20 |
> 20 ... 40 |
> 40 ... 100 |
Overpressure |
3 x FS |
3 x FS (≤ 12 bar) |
12 bar |
Accuracy, (1) (± % FS) |
≤ 0.5 / ≤ 0.25 |
≤ 0.5 / ≤ 0.25 |
≤ 0.5 / ≤ 0.25 |
Setting accuracy |
|
|
|
Zero point |
± 1mV |
± 1mV |
± 1mV |
Span |
± 2 % |
± 2 % |
± 2 % |
Thermal shift, (± % FS/°C) |
|
|
|
Zero point -5...50°C |
≤ 0.06 |
≤ 0.03 |
≤ 0.015 |
Span -5...50°C |
≤ 0.015 |
≤ 0.015 |
≤ 0.015 |
Long term stability, (2) |
≤ 0.2% FS / < 4 mbar |
≤ 0.1% FS / < 0.2% FS |
≤ 0.1% FS / < 0.2% FS |
(1) Best Straight Line (BSL) at ambient temperature
(2) 1 year (typ. / max.)
TYPICAL OUTPUT SIGNAL (MH2O)
|
10 ... 20 |
> 20 ... 40 |
> 40 ... 100 |
Output signal, (1), (mV) |
50 |
100 |
100 |
(1) At nominal pressure and 10 V DC
Technical specifications for active version, (1)
PRESSURE MEASURING RANGE (MH2O)
|
10 ... 20 |
> 20 ... 40 |
> 40 ... 100 |
Accuracy, (2) (± % FS) |
≤ 0.5 / ≤ 0.25 / ≤ 0.1 |
≤ 0.5 / ≤ 0.25 / ≤ 0.1 |
≤ 0.5 / ≤ 0.25 / ≤ 0.1 |
Thermal error (± % FS/°C) |
|
|
|
-5 ... 50°C compensated |
0.075 |
0.045 |
0.03 |
Thermal shift, (± % FS/°C) |
|
|
|
Zero point -5...50°C |
≤ 0.06 |
≤ 0.03 |
≤ 0.015 |
Span -5...50°C |
≤ 0.015 |
≤ 0.015 |
≤ 0.015 |
(1) For further specifications see ATM/N
(2) Zero based accuracy according to DIN16086, incl. hysteresis and repeatability at ambient temperature
|
Ordering information
(2)
|
|
|
X. |
XX XX. |
XXXX. |
XX. |
XXX |
Type |
|
|
|
|
|
|
|
|
MTM/N 10 |
15 |
|
|
|
|
|
Pressure type |
|
|
|
|
|
|
|
|
Gauge |
|
1 |
|
|
|
|
|
Absolute (vacuum) |
|
2 |
|
|
|
|
Pressure measuring range (1) |
|
|
|
|
|
|
|
|
1 bar ... 10 bar |
|
XX |
|
|
|
|
|
Offset, special adjustment |
|
99 |
|
|
|
|
Model |
|
|
|
|
|
|
|
|
Standart passive, with mV output signal, (Fig. 1/2) |
|
0 |
|
|
|
|
|
Active, with analog amplifier, (Fig. 3) |
|
2 |
|
|
|
|
Electrical connection |
|
|
|
|
|
|
|
|
PUR cable, black IP 68, (2) |
|
0 |
|
|
|
|
|
PE cable, black, IP 68, (2) |
|
1 |
|
|
|
|
Version |
|
|
|
|
|
|
|
|
Closed, (Fig. 1) |
|
|
55 |
|
|
|
|
Open, (Fig. 2) |
|
|
56 |
|
|
|
Connecting cable (4) |
|
|
|
|
|
|
|
|
PE cable |
|
|
|
|
|
1 |
|
PUR cable |
|
|
|
|
|
0 |
Output signal |
|
|
|
|
|
|
|
|
0...mV (specified by the customer), (Fig. 1 / 2), (8) |
|
|
|
13 |
|
|
|
0...100 mV |
|
|
|
14 |
|
|
|
0...5 VDC, (Fig. 3), (9) |
|
|
|
46 |
|
|
|
0...10 VDC, (Fig. 3), (9) |
|
|
|
47 |
|
|
|
0...20 mA, (Fig. 3), (9) |
|
|
|
00 |
|
|
|
4...20 mA, (Fig. 3), (9) |
|
|
|
05 |
|
|
Accuracy |
|
|
|
|
|
|
|
|
≤ ± 0.5 % FS, BSL (passive version only) (7), (8) |
|
|
|
|
0 |
|
|
≤ ± 0.5 % FS, (with option seperate electronic) (9) |
|
|
|
|
0 |
|
|
≤ ± 0.25 % FS, BSL (passive version only) ( (7), (8) |
|
|
|
|
1 |
|
|
≤ ± 0.25 % FS, (with option seperate electronic) (9) |
|
|
|
|
1 |
|
|
≤ ± 0.1 % FS, (with option seperate electronic) (9) |
|
|
|
|
2 |
|
Temperature range |
|
|
|
|
|
|
|
|
-5 ... 50 °C compensated
(allowed process temperature: -5...50 °C) |
|
|
|
|
4 |
|
Option |
|
|
|
|
|
|
|
|
Seals: NBR (Standard) |
|
|
|
|
|
R |
|
Titanium |
|
|
|
|
|
K |
|
Humidity filter element for gauge versions
(only for PUR and PE cable) |
|
|
|
|
|
Z |
(2) Please specify the required cable length and medium
(7) BSL: Best Straight Line
(8) Passive version
(9) Active version
|
|